VF-054SAHL

Print
Popis
Product Type Plasma Gel
Catalogue Number VF-054SAHL
Nominal Volume ml 5
Draw Volume ml 3.5
Additive Gel + 52.5 USP. U. Lithium Heparin
Concentration of Additive USP. U./ml 15
Dimensions - Length mm 75
Dimensions - Diameter mm 13
Stopper Colour  Light green
Cap Colour  Light green
Shelf Life month 24
Centrifugation Rate G/min 1200/10
Unit Box pcs 100
Shipping Carton pcs 12x100
Measures Unit Box LxWxH 284x85x69

Instructions for Use:

  • Immediately after sampling, the VF-054SAHL tube needs to be mixed gently and carefully (inversion 5 times).
  • Centrifuge the tube at room temperature, during at least 10 min. with a relative centrifugal force of 1200 G (+/-100 G).
    For optimal barrier formation, a swing-out rotor is recommended.
    Higher centrifugation force 4000 G (with stopper) or up to 13 000 G (without stopper) can be applied to get better platelet clearance.
    Lower centrifugation force might result in an incomplete gel barrier.
  • After complete separation with the gel barrier in place storage requirements depend on the stability of the analytes and are equivalent to storage requirements for decanted plasma samples.